Safety and biological aspects of present techniques of haemodialysis
Umeå University, Faculty of Medicine, Public Health and Clinical Medicine.
2006 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Introduction: Haemodialysis (HD) is a treatment in which blood from the patient is lead through a tubing system into a dialysis device in a extracorporeal circuit. This circuit contains semipermeable membranes (dialyzer). Blood with uraemic toxins flows on one side, and a salt solution flows on the other side. The salt solution flushes away waste products that have passed the membrane by diffusion or convection through small pores. From there the blood returns to the patient through a tubing system that contains an air-trap and a sensor to avoid air contamination in the blood. Besides air contamination, this treatment is burdened with safety problems such as biocompatibility, electrical safety and mechanical safety. The aim of this thesis was to investigate the safety issues in haemodialysis devices regarding leakage current and air contamination during standard procedures and simulated fault conditions. Does the dialysis device constitute a risk for the patient?

Methods: To determine the extent of leakage current in HD machines, measurements at the filter-coupling site were performed in vitro according to the safety standard, IEC 601-1, in 5 types of dialysis machines. To determine, in vitro, to what extent blood and priming fluid allowed leakage current to pass to the patient, leakage current were also measured in the blood lines. The blood line was filled with blood from donors or priming fluid in eight different runs. To determine if leakage current could influence biocompatibility, a Fresenius 2008C dialysis machine and 8 hemophan dialyzers were used. Blood lines contained about 400 ml heparinized blood from each of 8 different donors (in vitro). C3d was measured, in vitro, before start of a simulated dialysis and at 15, 30, 45 and 60 min. during standard dialysis procedure. Then 1.5 mA current was switched on and additional samples were drawn at 75 and 90 min. Some patients need a central dialysis catheter (CDC) for access, placed close to or within the heart. To analyze if leakage current during standard HD would influence the ECG, patients with CDC or with AV-fistula as access were investigated. To analyse if air contamination could occur without activating security alarms in the dialysis device, various modes of in vitro dialysis settings were studied, some using a dextran solution to mimic blood viscosity. Besides visual inspection an ultrasound detector for microemboli and microbubbles was also used.

Results: The data showed leakage current at the filter coupling site that was significantly higher for some devices than for others. The leakage current could pass through blood and priming fluid. It exceeded the cardiac floating (CF)-safety limit (<50μA) at the top of the CDC using the test mains on applied part for saline (median 1008μA), for blood (median 610μA) and for a single fault condition using saline (median 68 μA) or blood (47 μA). The leakage current experiments showed that complement activation worsened as the leakage current increased. During standard dialysis arrhythmia could occur. Microbubbles were visible at the bottom of the air-trap and bubbles could pass the air-trap towards the venous line without triggering the alarm. During recirculation, several ml of air could be collected in an intermediate bag after the venous line. Ultrasound showed the presence of bubbles of sizes 2.5-50 μm as well as more than 50 μm silently passing to the venous line in all runs performed.

In conclusion, the data showed that a leakage current in HD devices can be high enough to be a safety risk for the patient. This risk is greater if a single fault arises in the dialysis machine or another device connected to the same patient, or during mains contact to the patient. Then the current flow may be high enough to cause arrhythmia for the patient, especially when using a CDC. There is also reason for concern that micro bubble transmission may occur without inducing an alarm. These factors need to be looked over to improve safety regulations and optimize HD treatment and service schedules.

Place, publisher, year, edition, pages
2006. , p. 99
Series
Umeå University medical dissertations, ISSN 0346-6612 ; 1071
Keywords [en]
haemodialysis, safety, leakage current, central dialysis catheter, microbubbles, microemboli, air contamination
Research subject
transfusion medicine
Identifiers
URN: urn:nbn:se:umu:diva-941ISBN: 91-7264-226-2 (print)OAI: oai:DiVA.org:umu-941DiVA, id: diva2:145138
Public defence
2006-12-08, Sal B, 9 tr, Tandläkarhögskolan, NUS, Umeå, 13:00 (English)
Opponent
Available from: 2006-11-20 Created: 2006-11-20 Last updated: 2009-10-09Bibliographically approved
List of papers
1. Current leakage in hemodialysis machines may be a safety risk for patients.
Open this publication in new window or tab >>Current leakage in hemodialysis machines may be a safety risk for patients.
2000 (English)In: Artificial Organs, ISSN 0160-564X, E-ISSN 1525-1594, Vol. 24, no 12, p. 977-981Article in journal (Refereed) Published
Identifiers
urn:nbn:se:umu:diva-5577 (URN)10.1046/j.1525-1594.2000.06557.x (DOI)11121978 (PubMedID)
Available from: 2006-11-20 Created: 2006-11-20 Last updated: 2018-06-09Bibliographically approved
2. Blood lines conduct leakage current during haemodialysis: a potential safety risk during first failure, especially for patients with central dialysis catheter as access.
Open this publication in new window or tab >>Blood lines conduct leakage current during haemodialysis: a potential safety risk during first failure, especially for patients with central dialysis catheter as access.
2005 (English)In: Medical and Biological Engineering and Computing, ISSN 0140-0118, E-ISSN 1741-0444, Vol. 43, no 6, p. 731-738Article in journal (Refereed) Published
Keywords
Catheterization; Central Venous/*adverse effects, Electric Conductivity, Equipment Failure, Equipment Safety, Humans, Renal Dialysis/*instrumentation, Risk Assessment/methods
Identifiers
urn:nbn:se:umu:diva-14775 (URN)16594299 (PubMedID)
Available from: 2007-06-18 Created: 2007-06-18 Last updated: 2018-06-09Bibliographically approved
3. Electrical current leakage during hemodialysis may increase blood-membrane interaction
Open this publication in new window or tab >>Electrical current leakage during hemodialysis may increase blood-membrane interaction
2001 (English)In: International Journal of Artificial Organs, ISSN 0391-3988, E-ISSN 1724-6040, Vol. 24, no 3, p. 136-139Article in journal (Refereed) Published
Abstract [en]

During hemodialysis blood - membrane interaction causes complement activation. During dialysis there may be an electrical current leakage to the dialyzer, especially if there is a broken ground or a defect in another electrical device coupled to the patient.

This study investigated whether an electric current of 1.5 mA DC could alter blood membrane interaction as measured by changes in C3d in the blood. Such a high current leakage could occur because there is no protection in the dialysis machine (Class 1B) against auxiliary current leakage. Such a current could come from a defective external device in contact with the patient during hemodialysis.

Materials: A dialysis machine (Fresenius 2008C) with a filled blood-line system containing about 350 ml whole blood from each of 8 different donors was used in vitro. Each of the eight test-runs also contained 1000 U added heparin. The dialysis procedure was performed using hemophan membranes (GFS +12, Gambro) with bicarbonate and potassium 3.0 (D210, Gambro) as dialysate. Two electric poles were placed in the blood line, before and after the dialyzer (connected in parallel) and the ground was placed at entry and exit of the dialysate fluid coming from the machine to the dialysis filter. C3d was measured before the start of “dialysis” and at 15, 30, 45 and 60 min, during dialysis. Thereafter the 1.5 mA current was switched on and additional samples were drawn at 75 and 90 min. The mean C3d values were calculated. Paired non-parametric statistical analyses were performed.

Results: There was a significant and continuous increase in C3d as compared to the “predialysis” level. The increase during 0 to 30 minutes was greater than that from 30 to 60 minutes (p=0.018); the increase in C3d during 60 to 90 min, was greater than that from 30 to 60 min (p=0.018) and there was no difference between the 0 to 30 and the 60 to 90 min increases.

Conclusions: A current, used in this study, was able to induce a blood membrane interaction during in vitro dialysis. Even a weaker current leakage might have such adverse effects and similar interactions seem possible during regular dialysis depending on the extent of the leakage.

Place, publisher, year, edition, pages
Wichtig Editore Srl, 2001
Keywords
Hemodialysis, Current leakage, Biocompatibility, Complement activation, Membrane, Technical equipment
National Category
Biomedical Laboratory Science/Technology
Identifiers
urn:nbn:se:umu:diva-5579 (URN)10.1177/039139880102400304 (DOI)000167962100004 ()11314806 (PubMedID)2-s2.0-0035155241 (Scopus ID)
Available from: 2006-11-20 Created: 2006-11-20 Last updated: 2019-08-30Bibliographically approved
4. Measurement of cardiac rhythm in connection to haemodialysis with focus on a possible interference due to leakage current: a pilot study of patients in chronic haemodialysis
Open this publication in new window or tab >>Measurement of cardiac rhythm in connection to haemodialysis with focus on a possible interference due to leakage current: a pilot study of patients in chronic haemodialysis
Show others...
(English)Article in journal (Refereed) Submitted
Identifiers
urn:nbn:se:umu:diva-5580 (URN)
Available from: 2006-11-20 Created: 2006-11-20 Last updated: 2018-06-09Bibliographically approved
5. Air bubbles pass the security system of the dialysis device without alarming.
Open this publication in new window or tab >>Air bubbles pass the security system of the dialysis device without alarming.
Show others...
2007 (English)In: Artificial Organs, ISSN 0160-564X, E-ISSN 1525-1594, Vol. 31, no 2, p. 132-139Article in journal (Refereed) Published
Keywords
Embolism; Air/*diagnosis/*prevention & control, Equipment Failure Analysis, Filtration/instrumentation, Humans, Microbubbles, Renal Dialysis/*adverse effects
Identifiers
urn:nbn:se:umu:diva-8298 (URN)10.1111/j.1525-1594.2007.00352.x (DOI)17298402 (PubMedID)
Available from: 2008-03-06 Created: 2008-03-06 Last updated: 2018-06-09Bibliographically approved

Open Access in DiVA

fulltext(518 kB)7075 downloads
File information
File name FULLTEXT01.pdfFile size 518 kBChecksum SHA-1
735d3231b06ac368876c59a312fe08bb2912239b591762a498402f9b17919584f99d04ed
Type fulltextMimetype application/pdf

By organisation
Public Health and Clinical Medicine

Search outside of DiVA

GoogleGoogle Scholar
Total: 7075 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

isbn
urn-nbn

Altmetric score

isbn
urn-nbn
Total: 971 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf