Blood lines conduct leakage current during haemodialysis: a potential safety risk during first failure, especially for patients with central dialysis catheter as access.
Umeå universitet, Medicinska fakulteten, Institutionen för folkhälsa och klinisk medicin, Medicin.
Umeå universitet, Medicinska fakulteten, Institutionen för folkhälsa och klinisk medicin, Medicin.
2005 (engelsk)Inngår i: Medical and Biological Engineering and Computing, ISSN 0140-0118, E-ISSN 1741-0444, Vol. 43, nr 6, s. 731-738Artikkel i tidsskrift (Fagfellevurdert) Published
sted, utgiver, år, opplag, sider
2005. Vol. 43, nr 6, s. 731-738
Emneord [en]
Catheterization; Central Venous/*adverse effects, Electric Conductivity, Equipment Failure, Equipment Safety, Humans, Renal Dialysis/*instrumentation, Risk Assessment/methods
Identifikatorer
URN: urn:nbn:se:umu:diva-14775PubMedID: 16594299OAI: oai:DiVA.org:umu-14775DiVA, id: diva2:154447
Tilgjengelig fra: 2007-06-18 Laget: 2007-06-18 Sist oppdatert: 2018-06-09bibliografisk kontrollert
Inngår i avhandling
1. Safety and biological aspects of present techniques of haemodialysis
Åpne denne publikasjonen i ny fane eller vindu >>Safety and biological aspects of present techniques of haemodialysis
2006 (engelsk)Doktoravhandling, med artikler (Annet vitenskapelig)
Abstract [en]

Introduction: Haemodialysis (HD) is a treatment in which blood from the patient is lead through a tubing system into a dialysis device in a extracorporeal circuit. This circuit contains semipermeable membranes (dialyzer). Blood with uraemic toxins flows on one side, and a salt solution flows on the other side. The salt solution flushes away waste products that have passed the membrane by diffusion or convection through small pores. From there the blood returns to the patient through a tubing system that contains an air-trap and a sensor to avoid air contamination in the blood. Besides air contamination, this treatment is burdened with safety problems such as biocompatibility, electrical safety and mechanical safety. The aim of this thesis was to investigate the safety issues in haemodialysis devices regarding leakage current and air contamination during standard procedures and simulated fault conditions. Does the dialysis device constitute a risk for the patient?

Methods: To determine the extent of leakage current in HD machines, measurements at the filter-coupling site were performed in vitro according to the safety standard, IEC 601-1, in 5 types of dialysis machines. To determine, in vitro, to what extent blood and priming fluid allowed leakage current to pass to the patient, leakage current were also measured in the blood lines. The blood line was filled with blood from donors or priming fluid in eight different runs. To determine if leakage current could influence biocompatibility, a Fresenius 2008C dialysis machine and 8 hemophan dialyzers were used. Blood lines contained about 400 ml heparinized blood from each of 8 different donors (in vitro). C3d was measured, in vitro, before start of a simulated dialysis and at 15, 30, 45 and 60 min. during standard dialysis procedure. Then 1.5 mA current was switched on and additional samples were drawn at 75 and 90 min. Some patients need a central dialysis catheter (CDC) for access, placed close to or within the heart. To analyze if leakage current during standard HD would influence the ECG, patients with CDC or with AV-fistula as access were investigated. To analyse if air contamination could occur without activating security alarms in the dialysis device, various modes of in vitro dialysis settings were studied, some using a dextran solution to mimic blood viscosity. Besides visual inspection an ultrasound detector for microemboli and microbubbles was also used.

Results: The data showed leakage current at the filter coupling site that was significantly higher for some devices than for others. The leakage current could pass through blood and priming fluid. It exceeded the cardiac floating (CF)-safety limit (<50μA) at the top of the CDC using the test mains on applied part for saline (median 1008μA), for blood (median 610μA) and for a single fault condition using saline (median 68 μA) or blood (47 μA). The leakage current experiments showed that complement activation worsened as the leakage current increased. During standard dialysis arrhythmia could occur. Microbubbles were visible at the bottom of the air-trap and bubbles could pass the air-trap towards the venous line without triggering the alarm. During recirculation, several ml of air could be collected in an intermediate bag after the venous line. Ultrasound showed the presence of bubbles of sizes 2.5-50 μm as well as more than 50 μm silently passing to the venous line in all runs performed.

In conclusion, the data showed that a leakage current in HD devices can be high enough to be a safety risk for the patient. This risk is greater if a single fault arises in the dialysis machine or another device connected to the same patient, or during mains contact to the patient. Then the current flow may be high enough to cause arrhythmia for the patient, especially when using a CDC. There is also reason for concern that micro bubble transmission may occur without inducing an alarm. These factors need to be looked over to improve safety regulations and optimize HD treatment and service schedules.

Publisher
s. 99
Serie
Umeå University medical dissertations, ISSN 0346-6612 ; 1071
Emneord
haemodialysis, safety, leakage current, central dialysis catheter, microbubbles, microemboli, air contamination
Forskningsprogram
transfusionsmedicin
Identifikatorer
urn:nbn:se:umu:diva-941 (URN)91-7264-226-2 (ISBN)
Disputas
2006-12-08, Sal B, 9 tr, Tandläkarhögskolan, NUS, Umeå, 13:00 (engelsk)
Opponent
Tilgjengelig fra: 2006-11-20 Laget: 2006-11-20 Sist oppdatert: 2009-10-09bibliografisk kontrollert

Open Access i DiVA

Fulltekst mangler i DiVA

Andre lenker

PubMedFulltext

Personposter BETA

Jonsson, PerStegmayr, Bernd

Søk i DiVA

Av forfatter/redaktør
Jonsson, PerStegmayr, Bernd
Av organisasjonen
I samme tidsskrift
Medical and Biological Engineering and Computing

Søk utenfor DiVA

GoogleGoogle Scholar

pubmed
urn-nbn

Altmetric

pubmed
urn-nbn
Totalt: 186 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf