Evaluation of Strata NSC and Codman Hakim adjustable cerebrospinal fluid shunts and their corresponding antisiphon devices: laboratory investigation
Department of Paediatric Surgery, University Hospital, Uppsala.
Umeå universitet, Medicinska fakulteten, Institutionen för farmakologi och klinisk neurovetenskap, Neurokirurgi.
Umeå universitet, Medicinska fakulteten, Institutionen för farmakologi och klinisk neurovetenskap, Neurologi.ORCID-id: 0000-0001-6451-1940
2009 (engelsk)Inngår i: Journal of Neurosurgery: Pediatrics, ISSN 1933-0707, E-ISSN 1933-0715, Vol. 3, nr 3, s. 166-172Artikkel i tidsskrift (Fagfellevurdert) Published
Abstract [en]

OBJECT: The authors investigated and compared the in vitro characteristics of 2 CSF shunts, the Strata NSC and the Codman Hakim, and their corresponding antisiphon devices (ASDs).

METHODS: Six new CSF shunts and the corresponding ASDs for each model were tested in an automated, computerized experimental setup based on pressure regulation. Opening pressure accuracy, resistance, sensitivity to abdominal pressure, antisiphon effect, and the influence of different ASD positions were determined.

RESULTS: In general the shunts performed according to the manufacturers' specifications. However, at the lowest setting, the opening pressure of the Strata NSC was close to 0, and in the Codman Hakim shunt, it was higher than specified. The resistance in the Codman Hakim shunt (5.4 mm Hg/ml/min) was much higher than that in the Strata NSC (3.6 mm Hg/ml/min). Abdominal pressure affected opening pressure in both valves. Positioning the Strata ASD above or below the ventricular catheter tip resulted in higher and lower opening pressures, respectively, than when it was placed in line with the catheter. The positioning of the Codman Hakim ASD did not influence the opening pressure.

CONCLUSIONS: Both CSF shunts work properly, but at the lowest setting the opening pressure of the Strata NSC was near 0 and in the Codman Hakim it was twice the manufacturer's specifications. The resistance in the Strata NSC was below the normal physiological range, and in the Codman Hakim device it was in the lower range of normal. The ASD did not change the shunt characteristics in the lying position and therefore might not do so in children. If this is the case, then a shunt system with an integrated ASD could be implanted at the first shunt insertion, thus avoiding a second operation and the possibility of infection.

sted, utgiver, år, opplag, sider
American Association of Neurological Surgeons , 2009. Vol. 3, nr 3, s. 166-172
Emneord [en]
adjustable shunt, antisiphon device, hydrocephalus, in vitro test, normal pressure hydrocephalus
HSV kategori
Identifikatorer
URN: urn:nbn:se:umu:diva-23951ISI: 000263868300002PubMedID: 19338461OAI: oai:DiVA.org:umu-23951DiVA, id: diva2:225871
Tilgjengelig fra: 2009-06-29 Laget: 2009-06-29 Sist oppdatert: 2018-06-08bibliografisk kontrollert

Open Access i DiVA

Fulltekst mangler i DiVA

PubMed

Personposter BETA

Koskinen, Lars-Owe DMalm, JanEklund, Anders

Søk i DiVA

Av forfatter/redaktør
Koskinen, Lars-Owe DMalm, JanEklund, Anders
Av organisasjonen
I samme tidsskrift
Journal of Neurosurgery: Pediatrics

Søk utenfor DiVA

GoogleGoogle Scholar

pubmed
urn-nbn

Altmetric

pubmed
urn-nbn
Totalt: 631 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf