Evaluation of ultrasound Tissue Velocity Imaging: a phantom study of velocity estimation in skeletal muscle low-level contractions
Umeå universitet, Medicinska fakulteten, Institutionen för strålningsvetenskaper, Radiofysik.
2013 (Engelska)Ingår i: BMC Medical Imaging, ISSN 1471-2342, E-ISSN 1471-2342, Vol. 13, nr 16Artikel i tidskrift (Refereegranskat) Published
Abstract [en]

Background: Tissue Velocity Imaging (TVI) is an ultrasound based technique used for quantitative analysis of the cardiac function and has earlier been evaluated according to myocardial velocities. Recent years several studies have reported applying TVI in the analysis of skeletal muscles. Skeletal tissue velocities can be very low. In particular, when performing isometric contractions or contractions of low force level the velocities may be much lower compared to the myocardial tissue velocities. Methods: In this study TVI was evaluated for estimation of tissue velocities below the typical myocardial velocities. An in-house phantom was used to see how different PRF-settings affected the accuracy of the velocity estimations. Results: With phantom peak velocity at 0.03 cm/s the error ranged from 31% up to 313% with the different PRF-settings in this study. For the peak velocities at 0.17 cm/s and 0.26 cm/s there was no difference in error with tested PFR settings, it is kept approximately around 20%. Conclusions: The results from the present study showed that the PRF setting did not seem to affect the accuracy of the velocity estimation at tissue velocities above 0.17 cm/s. However at lower velocities (0.03 cm/s) the setting was crucial for the accuracy. The PRF should therefore preferable be reduced when the method is applied in low-level muscle contraction.

Ort, förlag, år, upplaga, sidor
2013. Vol. 13, nr 16
Nyckelord [en]
Tissue Velocity Imaging, Ultrasound, Skeletal muscle, Phantom evaluation, Pulse repetition frequency
Nationell ämneskategori
Radiologi och bildbehandling
Identifikatorer
URN: urn:nbn:se:umu:diva-79055DOI: 10.1186/1471-2342-13-16ISI: 000321369400001OAI: oai:DiVA.org:umu-79055DiVA, id: diva2:698867
Tillgänglig från: 2014-02-25 Skapad: 2013-08-06 Senast uppdaterad: 2018-06-08Bibliografiskt granskad

Open Access i DiVA

fulltext(1989 kB)294 nedladdningar
Filinformation
Filnamn FULLTEXT01.pdfFilstorlek 1989 kBChecksumma SHA-512
88f88785d5b95ae28863d6686b0f6414e3781d22e5bf029d9467671b5d688d996fa936686c6b7b7166dc11c99ebf6edd63d8870a4b442b3b226e99b8f31eec0e
Typ fulltextMimetyp application/pdf

Övriga länkar

Förlagets fulltext

Personposter BETA

Grönlund, Christer

Sök vidare i DiVA

Av författaren/redaktören
Grönlund, Christer
Av organisationen
Radiofysik
I samma tidskrift
BMC Medical Imaging
Radiologi och bildbehandling

Sök vidare utanför DiVA

GoogleGoogle Scholar
Totalt: 294 nedladdningar
Antalet nedladdningar är summan av nedladdningar för alla fulltexter. Det kan inkludera t.ex tidigare versioner som nu inte längre är tillgängliga.

doi
urn-nbn

Altmetricpoäng

doi
urn-nbn
Totalt: 190 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf