Show others and affiliations
2010 (English)In: JIM - Journal of Immunological Methods, ISSN 0022-1759, E-ISSN 1872-7905, Vol. 352, no 1-2, 23-31 p.Article in journal (Refereed) Published
Abstract [en]

This multi-center validation study identified the lack of preparation of accurate and consistent protein standards as the main reason for a poor inter-laboratory CV. This issue is also relevant to other protein biomarkers based on this type of assay and will need to be solved in order to achieve an acceptable level of analytical accuracy. The raw data of this study is available online.

Place, publisher, year, edition, pages
2010. Vol. 352, no 1-2, 23-31 p.
Keyword [en]
Neurofilament; NfL; NF-L; NEFL; Biomarker; Validation
URN: urn:nbn:se:umu:diva-41961DOI: 10.1016/j.jim.2009.09.014ISI: 000274102000003PubMedID: 19857497OAI: diva2:408192