Hydrodynamics of the CertasTM programmable valve for the treatment of hydrocephalus
Umeå University, Faculty of Medicine, Department of Radiation Sciences, Radiation Physics.
Umeå University, Faculty of Medicine, Department of Pharmacology and Clinical Neuroscience, Neurosurgery.
Show others and affiliations
2012 (English)In: Fluids and barriers of the CNS, ISSN 2045-8118, Vol. 9, no 1, 12- p.Article in journal (Refereed) Published
Abstract [en]

BACKGROUND: The new CertasTM shunt for the treatment of hydrocephalus has seven standard pressure settings that according to the manufacturer range from 36 to 238 mmH2O, and an additional "Virtual Off" setting with an opening pressure >400 mmH2O. Information on actual pressure response and reliability of shunt performance is important in clinical application, especially the "Virtual Off" setting as a non-surgical replacement for shunt ligation. The objective of this study was to evaluate the in-vitro hydrodynamic performance of the CertasTM shunt.

METHODS: Six new CertasTM shunts with proximal and distal catheters were tested with an automated, computerized test system that raised the pressure from zero to a maximum pressure and back to zero at each valve setting. Opening pressure and flow resistance were determined.

RESULTS: For settings 1-7 the measured opening pressure range was 26 to 247 mmH2O, and the mean change in opening pressure for a one-step adjustment was between 33 and 38 mmH2O. For setting 8 ("Virtual Off") the measured mean opening pressure was 494 +/- 34 mmH2O (range 451 to 556 mmH2O). The mean outflow resistance was 7.0 mmHg/ml/min (outflow conductance 17.9 ul/s/kPa).

CONCLUSIONS: The six shunts had similar characteristics and closely matched the manufacturer's specifications for opening pressure at settings 1-7. The opening pressure for the "Virtual Off" setting was nearly 500 mmH2O, which is 100 mmH2O higher than the manufacturer's specification of ">400" and should be functionally off for most patients with communicating hydrocephalus. Clinical studies are needed to evaluate if the CSF dynamic profile persists after implantation in patients.

Place, publisher, year, edition, pages
2012. Vol. 9, no 1, 12- p.
National Category
Medical Equipment Engineering
URN: urn:nbn:se:umu:diva-55836DOI: 10.1186/2045-8118-9-12PubMedID: 22643114OAI: diva2:530663
Available from: 2012-06-04 Created: 2012-06-04 Last updated: 2015-10-01Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Eklund, AndersKoskinen, Lars-Owe DMalm, Jan
By organisation
Radiation PhysicsNeurosurgeryDepartment of Pharmacology and Clinical Neuroscience
Medical Equipment Engineering

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 92 hits
ReferencesLink to record
Permanent link

Direct link