Construction of a Labview controlled pyrolysis unit for coupling to a Pyrola 85 pyrolysis chamber
Umeå University, Faculty of Science and Technology, Department of Chemistry.
Umeå University, Faculty of Science and Technology, Department of Chemistry.
2012 (English)Independent thesis Advanced level (degree of Master (One Year)), 20 credits / 30 HE creditsStudent thesis
Abstract [en]

Pyrolysis is the process of molecular decomposition in an inert environment using heat. It is possible to fragment large molecules, such as polymers, by pyrolysis and separate the fragments directly in a GC. This makes it possible to form complex sample fingerprints that can be used in various applications, for example in forensic science.

In this project, a malfunctioning Pyrola 85 pyrolysis unit was fixed by measuring the voltage signals from the photo diode during pyrolysis in a Labview program. With the Labview program, a partly manual filament temperature calibration procedure was developed, as replacement for the non-working automatic calibration procedure. The functioning pyrolysis unit was then installed on a GC/MS in the National University of Science, Ho Chi Minh City, Vietnam. Plant samples were analysed with the Pyrolysis-GC/MS system as a test of the system, resulting in pyrograms with a few identifiable peaks. However, both pyrolysis and GC/MS settings needs further investigation to optimize the analysis.

In Sweden, a Labview controlled Pyrolysis unit was constructed to investigate the possibility of improving certain functions of the original control unit. A proposed Labview program and circuit boards were constructed and partially evaluated. The accuracy and precision of the pulses from the Labview program was

Place, publisher, year, edition, pages
2012. , 32 p.
National Category
Analytical Chemistry
Identifiers
URN: urn:nbn:se:umu:diva-56549OAI: oai:DiVA.org:umu-56549DiVA: diva2:535725
External cooperation
National University of Science, Ho Chi Minh City, Vietnam
Uppsok
Physics, Chemistry, Mathematics
Available from: 2012-06-21 Created: 2012-06-20 Last updated: 2012-06-21Bibliographically approved

Open Access in DiVA

fulltext(931 kB)1043 downloads
File information
File name FULLTEXT01.pdfFile size 931 kBChecksum SHA-512
e77adf400933387465e4fc46dbdbada032a4e259224f0f91d1981c8babf0d3dcec74488f56d69e8792ba4fa314983e5a6f17adc0c362d7e4be3349eff02939c6
Type fulltextMimetype application/pdf

By organisation
Department of Chemistry
Analytical Chemistry

Search outside of DiVA

GoogleGoogle Scholar
Total: 1043 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

urn-nbn

Altmetric score

urn-nbn
Total: 303 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf