Microbubbles of air may occur in the organs of hemodialysis patients
Umeå University, Faculty of Medicine, Department of Public Health and Clinical Medicine, Medicine.
Umeå University, Faculty of Medicine, Department of Medical Biosciences, Pathology.
Umeå University, Faculty of Medicine, Department of Public Health and Clinical Medicine, Medicine.
Umeå University, Faculty of Medicine, Department of Public Health and Clinical Medicine, Medicine.
Show others and affiliations
2012 (English)In: ASAIO journal (1992), ISSN 1058-2916, E-ISSN 1538-943X, Vol. 58, no 2, 177-179 p.Article in journal (Refereed) Published
Abstract [en]

During hemodialysis (HD), blood that passes the dialysis device gets loaded with microbubbles (MB) of air that are returned to the patient without inducing an alarm. The aim with this study was to clarify if these signals are due to microembolies of air, clots, or artifacts, by histopathology of autopsy material of HD patients. These first results are from a patient on chronic HD. Due to pulmonary edema he was ultrafiltered. Within 30 minutes after the start, he suffered from a cardiac arrest and died. Autopsy verified the clinical findings. Microscopic investigation verified microembolies of air that were surrounded by fibrin in the lungs, brain, and heart. The study verified that MBs can enter the blood during HD and are trapped in the lungs. In addition, MBs pass the pulmonary capillaries and enter the arterial part of the body and are dispersed throughout the body. This can contribute to organ damage and be part of the poor prognoses seen in HD patients. Data support the importance to reduce MBs in the dialysis circuit.

Place, publisher, year, edition, pages
2012. Vol. 58, no 2, 177-179 p.
Keyword [en]
errors, laboratory medicine, preanalytical phase, quality, standardization
National Category
Urology and Nephrology
URN: urn:nbn:se:umu:diva-67084DOI: 10.1097/MAT.0b013e318245d0ddPubMedID: 22236622OAI: diva2:610576
Available from: 2013-03-12 Created: 2013-03-12 Last updated: 2013-04-25Bibliographically approved
In thesis
1. Presence of microemboli during haemodialysis and methods to reduce the exposure to microbubbles
Open this publication in new window or tab >>Presence of microemboli during haemodialysis and methods to reduce the exposure to microbubbles
2013 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Despite chronic dialysis treatment, patients with end stage renal disease undergoing maintenance haemodialysis (HD) remain at a substantially increased risk of morbidity. Previous reports using Doppler ultrasound (DU) during HD have revealed microembolic signals (ME) in the venous circulation.

In vitro studies confirm the emergence of microbubbles of air that may pass the security system of the HD circuit without triggering the alarm. The aim of this thesis was to elucidate the presence of ME during HD and examine methods that might reduce exposure to ME in vivo.

The first study utilized DU to verify the presence of ME in 40 patients during standard HD. Investigation within 30 minutes after the start of HD and just before the end of session revealed the presence of ME in the venous blood line during both phases. The air trap did not alert for the presence of ME. This indicated that ME may pass into the patient during the entire HD run.

Study 2 analyzed the presence of ME prior to start and during HD when measured at the AV-access and also carotid artery. A total of 54 patients were examined using DU as the investigative technique. ME increased significantly after start of HD in the AV-access, but also at the carotid artery site. These data indicated that ME can enter the body and even pass the lung barrier. The question arose if microbubbles of air are resorbed or may cause ischemic lesions in organs such as the brain.

Study 3 examined whether the amount of ME detected in the AV-access would change by using either a high or a low blood level in the venous air trap/chamber. This was a prospective, randomized and double-blind study of 20 HD patients who were their own controls. After 30 min of standard HD, measurement of ME with DU was performed for two minutes. The chamber setting was changed and after another 30 minutes a new recording was carried out for two minutes. Data showed that setting a high blood level significantly reduced the extent of ME that entered the patient. The results also indicated that ME consisted mainly of microbubbles.

In study 4, twenty patients were randomized in a cross-over setting of HD. Three options were used: a wet-stored dialyzer with high blood level (WH) and a dry-stored dialyzer using either a high (DH) or a low (DL) blood level in the venous chamber. The exposure of ME, detected by DU, was least when using mode WF, more with mode DH, and most with mode DL. There was a correlation between higher blood flow and more extensive exposure to ME.

Study 5 was an autopsy study of a chronic HD patient with the aim of searching for microbubbles deposited in organs. Microbubbles of gas were verified in the vessels of the lungs, brain and heart. By using a fluorescent stain of anti-fibrinogen it was verified that the microbubbles were covered by clots that had to be preformed before death occurred. This indicated that air microbubbles are not completely absorbed and could result in embolic deposition in the organs of HD patients.

In conclusion, these in vivo studies showed that ME pass the air trap without inducing an alarm and enter the venous blood line of the patient. The data confirmed the presence of ME in the AV-access and also in the carotid artery. Autopsy data of a deceased HD patient demonstrated the presence of microbubbles in the capillaries of the lungs, but also in the systemic circulation such as in the brain and the heart. A high blood level in the venous chamber and wet-stored dialyzer can reduce, but not eliminate the exposure to microbubbles for patients undergoing HD.

Place, publisher, year, edition, pages
Umeå: Umeå universitet, 2013. 106 p.
Umeå University medical dissertations, ISSN 0346-6612 ; 1564
Haemodialysis, in vivo, microemboli, microbubbles, dialyzer, air trap, venous chamber
National Category
Other Clinical Medicine Urology and Nephrology
urn:nbn:se:umu:diva-68674 (URN)978-91-7459-591-8 (ISBN)
Public defence
2013-05-16, Sal E04, Norrlands universitetssjukhus, Umeå, 09:00 (Swedish)
Available from: 2013-04-25 Created: 2013-04-22 Last updated: 2013-04-25Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Stegmayr, BerndBrännström, ThomasForsberg, UlfJonson, PerHultdin, Johan
By organisation
MedicinePathologyClinical chemistry
In the same journal
ASAIO journal (1992)
Urology and Nephrology

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 141 hits
ReferencesLink to record
Permanent link

Direct link