Optimal Exposure Time for Diagnosis of Approximal Caries Lesions on Digital X-rays
Umeå University, Faculty of Medicine, Department of Odontology, School of Dentistry.
Umeå University, Faculty of Medicine, Department of Odontology, School of Dentistry.
2014 (English)Independent thesis Advanced level (degree of Master (Two Years)), 20 credits / 30 HE creditsStudent thesis
Abstract [en]

In digital radiography the image can be enhanced and thereby an incorrect exposed radiograph can appear to have optimal brightness but with limited information and possibility to diagnose e.g. caries.

The primary aim of this study was to determine if different exposure times have an impact on the accuracy in diagnosis of approximal caries lesions in bitewing radiographs using a CMOS system. If so, what the optimal exposure time was.

A total of 40 extracted human teeth were selected and evaluated with a digital CMOS system. The teeth were exposed with different exposure times. CBCT was used to validate the presence and degree of approximal caries lesions on every individual tooth surface. The radiographs were evaluated by 11 general dentists and 4 specialists in oral and maxillofacial radiology.

The choice of exposure time did affect the diagnostic accuracy. A higher diagnostic accuracy was achieved when a longer exposure time (> 0.1 sec) was used when comparing kappa values and ROC curves among the observers. The radiologists had a higher diagnostic accuracy than the general dentists. A substantial variation in individual diagnostic accuracy between observers was found.

In conclusion; approximal caries was found to be easier to diagnose when the evaluated longer exposure times were used in the studied digital CMOS system. There was some decline in the longest exposures of the time span. However, when selecting optimal exposure time the patient's age and size should also be taken into consideration.

Place, publisher, year, edition, pages
2014.
National Category
Medical and Health Sciences
Identifiers
URN: urn:nbn:se:umu:diva-97855OAI: oai:DiVA.org:umu-97855DiVA: diva2:777112
Educational program
Dentistry Programme
Supervisors
Examiners
Available from: 2015-03-11 Created: 2015-01-08 Last updated: 2015-03-11Bibliographically approved

Open Access in DiVA

fulltext(505 kB)142 downloads
File information
File name FULLTEXT01.pdfFile size 505 kBChecksum SHA-512
8f932385ea0920d1889458eac04a42e363652a4aec5f6225300d104139f9944445250344cf13675bf6fa770d6dd45d7e78b9ea25369c581030ddf55971ac05be
Type fulltextMimetype application/pdf

By organisation
School of Dentistry
Medical and Health Sciences

Search outside of DiVA

GoogleGoogle Scholar
Total: 142 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

urn-nbn

Altmetric score

urn-nbn
Total: 107 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf