Change search
ReferencesLink to record
Permanent link

Direct link
A lab-on-a-chip for hypoxic patch clamp measurements combined with optical tweezers and spectroscopy-first investigations of single biological cells
Umeå University, Faculty of Medicine, Department of Radiation Sciences, Radiation Physics. Department of Computer Science, Electrical and Space Engineering, Luleå University of Technology; CMTF, Centre for Biomedical Engineering and Physics, Luleå and Umeå, Sweden; Department of Engineering Sciences and Mathematics, Luleå University of Technology.
2015 (English)In: Biomedical engineering online, ISSN 1475-925X, Vol. 14, 36Article in journal (Refereed) Published
Abstract [en]

The response and the reaction of the brain system to hypoxia is a vital research subject that requires special instrumentation. With this research subject in focus, a new multifunctional lab-on-a-chip (LOC) system with control over the oxygen content for studies on biological cells was developed. The chip was designed to incorporate the patch clamp technique, optical tweezers and absorption spectroscopy. The performance of the LOC was tested by a series of experiments. The oxygen content within the channels of the LOC was monitored by an oxygen sensor and verified by simultaneously studying the oxygenation state of chicken red blood cells (RBCs) with absorption spectra. The chicken RBCs were manipulated optically and steered in three dimensions towards a patch-clamp micropipette in a closed microfluidic channel. The oxygen level within the channels could be changed from a normoxic value of 18% O-2 to an anoxic value of 0.0-0.5% O-2. A time series of 3 experiments were performed, showing that the spectral transfer from the oxygenated to the deoxygenated state occurred after about 227 +/- 1 s and a fully developed deoxygenated spectrum was observed after 298 +/- 1 s, a mean value of 3 experiments. The tightness of the chamber to oxygen diffusion was verified by stopping the flow into the channel system while continuously recording absorption spectra showing an unchanged deoxygenated state during 5400 +/- 2 s. A transfer of the oxygenated absorption spectra was achieved after 426 +/- 1 s when exposing the cell to normoxic buffer. This showed the long time viability of the investigated cells. Successful patching and sealing were established on a trapped RBC and the whole-cell access (Ra) and membrane (Rm) resistances were measured to be 5.033 +/- 0.412 M Omega and 889.7 +/- 1.74 M Omega respectively.

Place, publisher, year, edition, pages
2015. Vol. 14, 36
Keyword [en]
Microfluidic system, Hypoxia, Patch clamp, Optical tweezers, Absorption spectroscopy, Oxygen sensor
National Category
Other Medical Biotechnology
URN: urn:nbn:se:umu:diva-103213DOI: 10.1186/s12938-015-0024-6ISI: 000353330300001PubMedID: 25907197OAI: diva2:813069
Available from: 2015-05-21 Created: 2015-05-18 Last updated: 2015-05-21Bibliographically approved

Open Access in DiVA

fulltext(3479 kB)78 downloads
File information
File name FULLTEXT01.pdfFile size 3479 kBChecksum SHA-512
Type fulltextMimetype application/pdf

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Lindahl, Olof A.
By organisation
Radiation Physics
In the same journal
Biomedical engineering online
Other Medical Biotechnology

Search outside of DiVA

GoogleGoogle Scholar
Total: 78 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 103 hits
ReferencesLink to record
Permanent link

Direct link