Accuracy of inverse treatment planning on substitute CT images derived from MR data for brain lesions
Umeå University, Faculty of Medicine, Department of Radiation Sciences.
Umeå University, Faculty of Medicine, Department of Radiation Sciences.
Umeå University, Faculty of Medicine, Department of Radiation Sciences.
Umeå University, Faculty of Medicine, Department of Radiation Sciences.
Show others and affiliations
2015 (English)In: Radiation Oncology, ISSN 1748-717X, Vol. 10, 13Article in journal (Refereed) Published
Abstract [en]

Background: In this pilot study we evaluated the performance of a substitute CT (s-CT) image derived from MR data of the brain, as a basis for optimization of intensity modulated rotational therapy, final dose calculation and derivation of reference images for patient positioning. Methods: S-CT images were created using a Gaussian mixture regression model on five patients previously treated with radiotherapy. Optimizations were compared using D-max, D-min, D-median and D-mean measures for the target volume and relevant risk structures. Final dose calculations were compared using gamma index with 1%/1 mm and 3%/3 mm acceptance criteria. 3D geometric evaluation was conducted using the DICE similarity coefficient for bony structures. 2D geometric comparison of digitally reconstructed radiographs (DRRs) was performed by manual delineation of relevant structures on the s-CT DRR that were transferred to the CT DRR and compared by visual inspection. Results: Differences for the target volumes in optimization comparisons were small in general, e.g. a mean difference in both D-min and D-max within similar to 0.3%. For the final dose calculation gamma evaluations, 100% of the voxels passed the 1%/1 mm criterion within the PTV. Within the entire external volume between 99.4% and 100% of the voxels passed the 3%/3 mm criterion. In the 3D geometric comparison, the DICE index varied between approximately 0.8-0.9, depending on the position in the skull. In the 2D DRR comparisons, no appreciable visual differences were found. Conclusions: Even though the present work involves a limited number of patients, the results provide a strong indication that optimization and dose calculation based on s-CT data is accurate regarding both geometry and dosimetry.

Place, publisher, year, edition, pages
2015. Vol. 10, 13
Keyword [en]
Radiotherapy, Treatment planning, MRI, Substitute CT, s-CT
National Category
Cancer and Oncology Radiology, Nuclear Medicine and Medical Imaging
URN: urn:nbn:se:umu:diva-102308DOI: 10.1186/s13014-014-0308-1ISI: 000350027400001PubMedID: 25575414OAI: diva2:825693
Available from: 2015-06-24 Created: 2015-04-23 Last updated: 2015-06-24Bibliographically approved

Open Access in DiVA

fulltext(1812 kB)60 downloads
File information
File name FULLTEXT01.pdfFile size 1812 kBChecksum SHA-512
Type fulltextMimetype application/pdf

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Jonsson, Joakim H.Akhtari, Mohammad M.Karlsson, Magnus G.Johansson, AdamAsklund, ThomasNyholm, Tufve
By organisation
Department of Radiation Sciences
In the same journal
Radiation Oncology
Cancer and OncologyRadiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 60 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 101 hits
ReferencesLink to record
Permanent link

Direct link