umu.sePublications
Change search
ReferencesLink to record
Permanent link

Direct link
Implant stability measurements for implants placed at the time of extraction: a cohort, prospective clinical trial
Umeå University, Faculty of Medicine, Department of Odontology, Oral and Maxillofacial Surgery. Department of Biomaterials, Institute for Surgical Sciences, University of Gothenburg.
Show others and affiliations
2005 (English)In: Journal of Periodontology, ISSN 0022-3492, E-ISSN 1943-3670, Vol. 76, no 3, 391-397 p.Article in journal (Refereed) PublishedText
Abstract [en]

Background: Studies indicate that implants placed at the time of extraction have high success rates. Implants must be stable at the time of insertion. Presently there are no data indicating the degree of implant stability when implants are placed at the time of extraction. This study evaluated changes in stability of implants from implant placement to abutment connection utilizing resonance frequency analysis (RFA). The unit of measurement was the international stability quotient (ISQ).

Methods: Prior to treatment, patients were given medical history and dental evaluations. Periapical and panogram radiographs were taken. Fifty-two patients requiring extraction of one or two teeth and implant placement immediately after extraction were enrolled in this study. Under conscious sedation and local anesthesia or local anesthesia alone, teeth were atraumatically removed and the extraction sockets were debrided. A total of 73 dental implants (57 in the maxilla, 16 in the mandible) were placed. Using a one-stage approach, all implants were placed within the patient's alveolar envelope and were never placed directly into extraction sockets. All implants were placed into contained extraction sites. Bone augmentation procedures were not performed. After implant insertion, the RFA electronic transducer was attached to the head of the implant with the retaining screw. The device was attached to a computer designed to register RFA scores in ISQ units. RFA measurements were taken at implant placement and abutment connection. Bone qualities, quantity, implant length and width as well as site of placement were recorded.

Results: The average interval between implant insertion and abutment connection was 5.6 months (SD 2.05). Two implants were lost between implant insertion and 1 year. At 2 to 3 years, the cumulative survival was 97.2%. Resonance frequency measurements at implant placement showed a mean primary stability of 62.0 (SE 1.1; range 43 to 83 ISQ) and a mean secondary stability after 1 year of 64.0 (SE 1.2; range 40 to 98 ISQ) for all implants. The increase was marginally significant (generalized estimating equation z-statistic = 1.79; P value = 0.07).

Conclusions: Implants placed at the time of extraction and inserted into native bone and not directly into extractions sockets have a high degree of initial stability as evidenced by RFA measurements. Implants with initial high ISQ levels revealed a slight drop in levels over time, while implants with levels lower than 60 had increases in levels between implant insertion and abutment connection. At 2 to 3 years the cumulative survival rate was 97.2%.

Place, publisher, year, edition, pages
Chicago: American academy of periodontology , 2005. Vol. 76, no 3, 391-397 p.
Keyword [en]
dental implantation, follow-up studies, international stability quotient, resonance frequency analysis, tooth extraction
National Category
Dentistry Surgery
Identifiers
URN: urn:nbn:se:umu:diva-120434DOI: 10.1902/jop.2005.76.3.391ISI: 000228660100011PubMedID: 15857073OAI: oai:DiVA.org:umu-120434DiVA: diva2:932856
Available from: 2016-06-02 Created: 2016-05-16 Last updated: 2016-06-02Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Sennerby, Lars
By organisation
Oral and Maxillofacial Surgery
In the same journal
Journal of Periodontology
DentistrySurgery

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 13 hits
ReferencesLink to record
Permanent link

Direct link