Upper limits of the photon fluence rate on CT detectors: case study on a commercial scanner
Show others and affiliations
2016 (English)In: Medical physics (Lancaster), ISSN 0094-2405, Vol. 43, no 7, 4398-4411 p.Article in journal (Refereed) PublishedText
Abstract [en]

Purpose: The highest photon fluence rate that a computed tomography (CT) detector must be able to measure is an important parameter. The authors calculate the maximum transmitted fluence rate in a commercial CT scanner as a function of patient size for standard head, chest, and abdomen protocols.

Methods: The authors scanned an anthropomorphic phantom (Kyoto Kagaku PBU-60) with the reference CT protocols provided by AAPM on a GE LightSpeed VCT scanner and noted the tube current applied with the tube current modulation (TCM) system. By rescaling this tube current using published measurements on the tube current modulation of a GE scanner [N. Keat, “CT scanner automatic exposure control systems,” MHRA Evaluation Report 05016, ImPACT, London, UK, 2005], the authors could estimate the tube current that these protocols would have resulted in for other patient sizes. An ECG gated chest protocol was also simulated. Using measured dose rate profiles along the bowtie filters, the authors simulated imaging of anonymized patient images with a range of sizes on a GE VCT scanner and calculated the maximum transmitted fluence rate. In addition, the 99th and the 95th percentiles of the transmitted fluence rate distribution behind the patient are calculated and the effect of omitting projection lines passing just below the skin line is investigated.

Results: The highest transmitted fluence rates on the detector for the AAPM reference protocols with centered patients are found for head images and for intermediate-sized chest images, both with a maximum of 3.4 ⋅ 108 mm−2 s−1, at 949 mm distance from the source. Miscentering the head by 50 mm downward increases the maximum transmitted fluence rate to 5.7 ⋅ 108 mm−2 s−1. The ECG gated chest protocol gives fluence rates up to 2.3 ⋅ 108 − 3.6 ⋅ 108 mm−2 s−1 depending on miscentering.

Conclusions: The fluence rate on a CT detector reaches 3 ⋅ 108 − 6 ⋅ 108 mm−2 s−1 in standard imaging protocols, with the highest rates occurring for ECG gated chest and miscentered head scans. These results will be useful to developers of CT detectors, in particular photon counting detectors.

Place, publisher, year, edition, pages
American Association of Physicists in Medicine , 2016. Vol. 43, no 7, 4398-4411 p.
Keyword [en]
fluence rate, count rate requirements, count rate problem, photon counting CT
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:umu:diva-124258DOI: 10.1118/1.4954008ISI: 000379171900044PubMedID: 27370155OAI: oai:DiVA.org:umu-124258DiVA: diva2:950662
Available from: 2016-08-01 Created: 2016-07-29 Last updated: 2016-08-01Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Andersson, Jonas
By organisation
Radiation Physics
In the same journal
Medical physics (Lancaster)
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 12 hits
ReferencesLink to record
Permanent link

Direct link