Radiation Dosimetry for CBCT – Quality Control and Applied Dosimetry
Umeå universitet, Medicinska fakulteten, Institutionen för strålningsvetenskaper, Radiofysik.
2015 (engelsk)Konferansepaper, Oral presentation only (Annet vitenskapelig)
Abstract [en]

The use of Cone Beam CT (CBCT) devices in health care has increased in recent years. Unfortunately, this trend has not been followed by a standardization of dose metrics for Quality Control (QC), or a necessary evolution of applied dosimetry for assessment of patient dose.

The European Federation of Organisations in Medical Physics (EFOMP) has drafted a report on the property of QC for CBCT devices. The report contains objective, practical and unifying methodology for QC of CBCT used in oral radiology, radiotherapy, interventional radiology and guided surgery. The dose metrics discussed include the Computed Tomography Dose Index (CTDI), Kerma-Area Product (KAP) and detector incident air kerma. The report concludes that KAP-meters are preferable for QC as long as they can be mounted on the X-ray tube housing. Otherwise measurements of detector incident kerma seem to offer a practical solution for QC.

The European Radiation Dosimetry Group (EURADOS) are preparing a literature review on patient dose from various applications of CBCT, which will in part be included in the EFOMP report. Most studies on patient dose from CBCT examinations and interventional procedures are based on thermoluminescent dosimeter (TLD) measurements in Rando phantoms. While helpful to the community, these studies yield substantial uncertainty when applied in the clinical reality of medical physicists working with justification and optimization. Applied dosimetry for CBCT needs to evolve and adopt recent theoretical advances to allow robust estimates of patient dose.

sted, utgiver, år, opplag, sider
2015.
HSV kategori
Forskningsprogram
radiofysik
Identifikatorer
URN: urn:nbn:se:umu:diva-113686OAI: oai:DiVA.org:umu-113686DiVA, id: diva2:888227
Konferanse
17th CT users group meeting, Childon, UK, December 15, 2015
Tilgjengelig fra: 2015-12-22 Laget: 2015-12-22 Sist oppdatert: 2021-08-12bibliografisk kontrollert

Open Access i DiVA

Fulltekst mangler i DiVA

Person

Andersson, Jonas

Søk i DiVA

Av forfatter/redaktør
Andersson, Jonas
Av organisasjonen

Søk utenfor DiVA

GoogleGoogle Scholar

urn-nbn

Altmetric

urn-nbn
Totalt: 1364 treff
RefereraExporteraLink to record
Permanent link

Direct link
Referera
Referensformat
  • apa
  • ieee
  • vancouver
  • Annet format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annet språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf