Sensitivity improvements of a resonance-based tactile sensor.
Umeå University, Faculty of Medicine, Department of Radiation Sciences.
2017 (English)In: Journal of Medical Engineering & Technology, ISSN 0309-1902, E-ISSN 1464-522X, Vol. 41, no 2, p. 131-140Article in journal (Refereed) Published
Abstract [en]

Resonance-based contact-impedance measurement refers to the application of resonance sensors based on the measurement of the changes in the resonance curve of an ultrasonic resonator in contact with a surface. The advantage of the resonance sensor is that it is very sensitive to small changes in the contact impedance. A sensitive micro tactile sensor (MTS) was developed, which measured the elasticity of soft living tissues at the single-cell level. In the present paper, we studied the method of improving the touch and stiffness sensitivity of the MTS. First, the dependence of touch sensitivity in relation to the resonator length was studied by calculating the sensitivity coefficient at each length ranging from 9 to 40 mm. The highest touch sensitivity was obtained with a 30-mm-long glass needle driven at a resonance frequency of 100 kHz. Next, the numerical calculation of contact impedance showed that the highest stiffness sensitivity was achieved when the driving frequency was 100 kHz and the contact-tip diameter of the MTS was 10 μm. The theoretical model was then confirmed experimentally using a phase-locked-loop-based digital feedback oscillation circuit. It was found that the developed MTS, whose resonant frequency was 97.030 kHz, performed with the highest sensitivity of 53.2 × 10(6) Hz/N at the driving frequency of 97.986 kHz, i.e. the highest sensitivity was achieved at 956 Hz above the resonant frequency.

Place, publisher, year, edition, pages
2017. Vol. 41, no 2, p. 131-140
Keywords [en]
Resonance-based sensor, Young’s modulus, contact impedance, stiffness sensitivity, tactile
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:umu:diva-133977DOI: 10.1080/03091902.2016.1239278PubMedID: 27701917Scopus ID: 2-s2.0-84990201247OAI: oai:DiVA.org:umu-133977DiVA, id: diva2:1090327
Available from: 2017-04-24 Created: 2017-04-24 Last updated: 2023-03-24Bibliographically approved

Open Access in DiVA

No full text in DiVA

Other links

Publisher's full textPubMedScopus

Authority records

Lindahl, Olof A

Search in DiVA

By author/editor
Lindahl, Olof A
By organisation
Department of Radiation Sciences
In the same journal
Journal of Medical Engineering & Technology
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 415 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf