Assessment of gas molar density by gas modulation refractometry: A review of its basic operating principles and extraordinary performance
Umeå universitet, Teknisk-naturvetenskapliga fakulteten, Institutionen för fysik.ORCID-id: 0000-0002-8580-9700
Umeå universitet, Teknisk-naturvetenskapliga fakulteten, Institutionen för fysik.ORCID-id: 0000-0001-5790-2185
Umeå universitet, Teknisk-naturvetenskapliga fakulteten, Institutionen för fysik. Measurement Science and Technology, RISE Research Institutes of Sweden, Borås, Sweden.ORCID-id: 0000-0001-6824-3111
Umeå universitet, Teknisk-naturvetenskapliga fakulteten, Institutionen för fysik.ORCID-id: 0000-0002-3261-9903
Visa övriga samt affilieringar
2021 (Engelska)Ingår i: Spectrochimica Acta Part B - Atomic Spectroscopy, ISSN 0584-8547, E-ISSN 1873-3565, Vol. 179, artikel-id 106121Artikel, forskningsöversikt (Refereegranskat) Published
Abstract [en]

A technique for high-precision and high-accuracy assessment of both gas molar (and number) density and pressure, Gas Modulation Refractometry (GAMOR), is presented. The technique achieves its properties by assessing refractivity as a shift of a directly measurable beat frequency by use of Fabry-Perot cavity (FPC) based refractometry utilizing the Pound-Drever-Hall laser locking technique. Conventional FPC-based refractometry is, however, often limited by fluctuations and drifts of the FPC. GAMOR remedies this by an additional utilization of a gas modulation methodology, built upon a repeated filling and evacuation of the measurement cavity together with an interpolation of the empty cavity responses. The procedure has demonstrated an ability to reduce the influence of drifts in a non-temperature stabilized dual-FPC (DFPC)-based refractometry system, when assessing pressure, by more than three orders of magnitude. When applied to a DFPC system with active temperature stabilization, it has demonstrated, for assessment of pressure of N2 at 4304 Pa at room temperature, which corresponds to a gas molar density of 1.7 × 10−6 mol/cm3, a sub-0.1 ppm precision (i.e. a resolution of 0.34 mPa). It is claimed that the ability to assess gas molar density is at least as good as so far has been demonstrated for pressure (i.e. for the molar density addressed, a resolution of at least 1.2 × 10−13 mol/cm3). It has recently been argued that the methodology should be capable of providing an accuracy that is in the low ppm range. These levels of precision and accuracy are unprecedented among laser-based techniques for detection of atomic and molecular species. Since the molar polarizability of He can be calculated by ab initio quantum mechanical calculations with sub-ppm accuracy, it can also be used as a primary or semi-primary standard of both gas molar (and number) density and pressure.

Ort, förlag, år, upplaga, sidor
Elsevier, 2021. Vol. 179, artikel-id 106121
Nyckelord [en]
Accuracy, Fabry-Perot cavity, Gas modulation refractometry (GAMOR), Gas molar (or number) density, Precision
Nationell ämneskategori
Annan elektroteknik och elektronik
Identifikatorer
URN: urn:nbn:se:umu:diva-182488DOI: 10.1016/j.sab.2021.106121ISI: 000631868700025Scopus ID: 2-s2.0-85103981287OAI: oai:DiVA.org:umu-182488DiVA, id: diva2:1546722
Forskningsfinansiär
Vetenskapsrådet, 621-2015-04374Kempestiftelserna, 1823Kempestiftelserna, U12Vinnova, 2017-05013Vinnova, 2018-04570Vinnova, 2019-05029EU, Horisont 2020Tillgänglig från: 2021-04-23 Skapad: 2021-04-23 Senast uppdaterad: 2023-09-05Bibliografiskt granskad

Open Access i DiVA

fulltext(4601 kB)224 nedladdningar
Filinformation
Filnamn FULLTEXT01.pdfFilstorlek 4601 kBChecksumma SHA-512
bf515c9680230a464e09b8eaf618fb892423a7bd29a074bdf6c67cd51bbc775c7751acdc0c91b9c40053dc72869ff946f50ee9b58fb980733820103fc5f12d66
Typ fulltextMimetyp application/pdf

Övriga länkar

Förlagets fulltextScopus

Person

Axner, OveSilander, IsakForssén, ClaytonZakrisson, Johan

Sök vidare i DiVA

Av författaren/redaktören
Axner, OveSilander, IsakForssén, ClaytonZakrisson, Johan
Av organisationen
Institutionen för fysik
I samma tidskrift
Spectrochimica Acta Part B - Atomic Spectroscopy
Annan elektroteknik och elektronik

Sök vidare utanför DiVA

GoogleGoogle Scholar
Totalt: 225 nedladdningar
Antalet nedladdningar är summan av nedladdningar för alla fulltexter. Det kan inkludera t.ex tidigare versioner som nu inte längre är tillgängliga.

doi
urn-nbn

Altmetricpoäng

doi
urn-nbn
Totalt: 455 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • vancouver
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf